SSI Technologies – Application Note PS-AN2
MediaSensor? Absolute, Sealed, Gage Pressure Sensors
Family Product Overview
Typical Connections
MediaSensor ? 4 -20mA Output connections:
The following torque limits should be used when mounting
the MediaSensor ? pressure port.
1) Connect the Power Lead (Red) to the + terminal of
the supply voltage.
2) Connect the Return Lead (White) to the + terminal
Straight Thread w/O-Ring:
Recommended Torque
of the current measuring device
High Pressure
(> 750
300 in lb
3) Connect the – terminal of the current measuring
PSI)
All others with out
Port types T, U, Y
150 in lb
device to the – terminal of the supply voltage.
MediaSensor ? Voltage Output connections:
Parts with Ports T,U, Y
NPTF Thread:
120 in lb
2 T.F.F.T. (Turns From Finger Tight)
1)
2)
Connect the Power Lead (Red) to the + terminal of
the supply voltage.
Connect the Ground Lead (Black) and the –
terminal of the supply voltage to – input of your
voltage measurement equipment.
SSI TECHNOLOGIES, INC.
Controls Division
2643 W. Court Street
Janesville, WI 53548-5011
3)
Connect the Vout Lead (White) to the + input of
your voltage measurement equipment.
Copyright January 12, 2009
SSI Technologies Inc.
All Rights Reserved
Revision 3
Phone: (608)758-1500 Fax: (608) 758-2491
www.ssitechnologies.com
Page 4
RoHS